Fluke Calibration 5080A High Compliance Multi-Product Calibrator

  • Fluke Calibration 5080A High Compliance Multi-Product CalibratorFluke Calibration 5080A High Compliance Multi-Product CalibratorCalibration solutions for your analog and digital workload High compliance Protection circuitry Calibrates a wide workload, including analog meters and 3.5 and 4.5 digits DMMs The 5080A Multi-Product Calibrator calibrates your analog and digital workload accurately and economically....



    Add to quote

    Fluke Calibration 5080A High Compliance Multi-Product Calibrator

    Fluke Calibration 5080A High Compliance Multi-Product Calibrator

    Calibration solutions for your analog and digital workload

    • High compliance
    • Protection circuitry
    • Calibrates a wide workload, including analog meters and 3.5 and 4.5 digits DMMs

    The 5080A Multi-Product Calibrator calibrates your analog and digital workload accurately and economically. Its high voltage and current compliance makes analog workload calibration easy and precise. And built-in protection circuitry protects it against damaging input voltages.

    This easy-to-use multifunction calibrator can handle a wide meter calibration workload that includes:

    • Analog meters
    • Panel meters
    • Digital multimeters
    • Watt meters
    • Clamp meters (with coil accessory)
    • Megohm meters (optional)
    • Oscilloscopes to 200 MHz (optional)
    • …and more

    Versatile software applications enable you to record paperless results, and more. Best of all, the 5080A Multi-Product Calibrator offers this versatile functionality at a price that fits nicely into your budget.

     

    Technical Specifications
    Voltage DC
      Range: 0 to ±1020V
      Best 1 Year Specification: ±0.01% of setting
    Current DC
      Range: 0 to ±20.5A
      Best 1 Year Specification: ±0.05% of setting
    Resistance
      Range: 0 to 190 MO, 19 values in x1 and x1.9
      Best 1 Year Specification: ±0.025% of setting
    Voltage AC
      Range: 1 mV to 1020V
    45Hz to 1 kHz, Sine
      Best 1 Year Specification: ±0.1% of setting
    Current AC
      Range:29 mA to 20.5A
    45 Hz to 1 kHz, Sine
      Best 1 Year Specification: ±0.1% of setting
    DC Power
      Range:10.9 mW to 20.5 kW
      Best 1 Year Specification: ±0.11% of setting
    AC Power
      Range: 10.9 mW to 20.5 kW
    45-65 Hz, PF=1
      Best 1 Year Specification: ±0.36% of setting
    Phase
      Range: 0 to ± 179.99°
    45 Hz to 1 kHz
      Best 1 Year Specification: ±0.25° of setting
    Frequency
      Range: 45 Hz to 1 kHz
      Best 1 Year Specification: ±50 ppm of setting
    General Specifications
    Warm up Time
      Twice the time since last warmed up, to a maximum of 30 minutes
    Settling Time
      Less than 5 seconds for all functions and ranges except as noted
    Standard Interfaces
      RS-232 and Ethernet
    Temperature Performance
    Operating: 0°C to 50°C
    Calibration (tcal):15°C to 35°C
    Storage:-20°C to +70°C
    Temperature Coefficient
      Temperature coefficient for temperatures outside tcal ±5 °C is 10 % of the stated specification per °C for temperatures in the range of 0 °C to 35 °C. Above 35 °C, the temperature coefficient is 20 % of the stated specification per °C.
    Relative Humidity
    Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C
    Storage:<95%, non-condensing
    Altitude
    Operating: 2,000m (6,500 ft) maximum
    Non-operating: 12,200m (40,000 ft) maximum
    Safety
    Meets EN. 61010-1:2001, CAN/CSA-C22.2 No. 61010-1-04, UL 61010-1: 2004
    Analog Low Isolation
    20V
    EMC
    Meets EN613261:2006.
    Line Power
    Line Voltage (selectable): 100V, 120V, 220V, 240V
    Line Frequency: 47 to 63 Hz
    Line Voltage Variation: + 10% about line voltage setting
    Power Consumption
    600 VA
    Calibration Documentation Report of calibration with data included; 17025 accredited report optional
    Dimensions
    Height:19.3 cm (7.6 in)
    Width:43.2 cm (17 in), 44.3 cm (17.5 in) including handles
    Depth:53.8 cm (21.2 in)
    Weight (without options)
    22 kg (48 lbs.)
    Absolute Uncertainty Definition
    The specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. It is not necessary to add anything to determine the total specification for the temperature range indicated.
    Specification Confidence Interval
    99%

    You can use this tab for adding shipping details, return policies and information or measurements etc.