Fluke Calibration 5322A Electrical Safety Tester Calibrator

  • Fluke Calibration 5322A Electrical Safety Tester CalibratorThe 5322A Electrical Tester Calibrator helps calibration technicians comply with new regulatory standards up to four times faster than with traditional manual, multiple-product methods.The 5322A replaces the 5320A Electrical Tester Calibrator. The 5322A calibrator facilitates compliance with exacting international standards such as the United Kingdom’s BS7671 17th Edition,...



    Add to quote

    Fluke Calibration 5322A Electrical Safety Tester Calibrator

    The 5322A Electrical Tester Calibrator helps calibration technicians comply with new regulatory standards up to four times faster than with traditional manual, multiple-product methods.The 5322A replaces the 5320A Electrical Tester Calibrator.
     
    The 5322A calibrator facilitates compliance with exacting international standards such as the United Kingdom’s BS7671 17th Edition, IEC/EN Standards, Australia and New Zealand’s AS/NZS 3000 and Chinese verification/calibration regulations for various electrical testers.
     
    The 5322A combines many functions into a single instrument, replacing discrete resistors, decade boxes and other custom solutions commonly used to calibrate electrical testers. This single-box solution speeds and simplifies calibration because users only need to learn, operate and maintain one calibrator rather than multiple instruments. And while it’s difficult to automate multiple instruments, the 5322A can be automated with MET/CAL® Calibration Software—further increasing speed and throughput.

    Calibrate all major types of electrical safety testers with just one calibrator. The 5322A calibrates all major categories of electrical safety testers. The benefits of this calibrator are best described by the key functionality it brings to calibrating the individual workloads.

    Broad workload coverage

    The 5322A calibrates a broad range of equipment, including:

    • Hipot testers
    • Insulation resistance testers (megohm-meters) including older analog testers
    • Loop/line impedance testers
    • Continuity testers
    • Earth resistance testers
    • Ground bond testers
    • Leakage current testers
    • Circuit breaker testers (RCD/GFCI)
    • Multifunction installation testers
    • Portable appliance testers (PATs)

     

    General specifications

    Specifications confidence level 99%
    Specifications interval 1 year
    Power line

    115/230 V ac (50/60 Hz) +10 % / -14 %, with the maximum voltage difference between neutral and protective earth not exceeding 15 V. Operation with power line between -10 % and -14 % has limitations in burden current for voltage outputs. See AC/DC Voltage Calibrator (VLC option) below.

    Power consumption 1250 VA maximum
    Fuse protection

    AC mains input: 2 A, 250 V for 230 V, time delay (T2L250 V – 5 mm x 20 mm) 4 A, 250 V for 115 V, time delay (T4L250 V – 5 mm x 20 mm)

    RCD input: 3.15 A, 250 V, fast (F3.15H250 V – 5 mm x 20 mm)

    Meter amps (A) input: 20 A, 500 V, time delay (F20H500 V – 6.3 mm x 32 mm)

    Loop/Line impedance input: 4 A, 500 V, time delay (T4H500 V – 6.3 mm x 32 mm)

    Leakage current input: 100 mA, 250 V, fast (F100 mL250 V – 5 mm x 20 mm)

    Environment
    Warm-up time 15 minutes
    Temperature performance

    Operating temperature: 18 °C to 28 °C

    Calibration temperature (tcal): 23 °C

    Temperature coefficient for temperatures outside of tcal ± 5 °C is 0.1 / °C of the specification, for temperatures between 5 °C to 40 °C.

    Storage temperature : -10 °C to 50 °C

    Storage recovery time: Typically <24 hours at operating temperature

    Relative humidity (operating) <80 % to 28 °C (resistance outputs >10 GΩ specified for <70 % to 28 °C)
    Relative humidity (storage) 90 % non-condensing 0 °C to 50 °C
    Altitude

    Operating: 3050 m (10 000 ft)

    Storage: 12 200 m (40 000 ft)

    Dimensions and weight
    Dimensions 430 mm x 555 mm x 170 mm (16.9 in x 21.8 in x 6.7 in)
    Weight 20 kg (44.1 lb)
    Compliance
    Safety

    Mains: IEC 61010-1: Overvoltage category II, pollution degree 2

    Measurement: IEC 61010-2-030: 5000 V (not category rated)

    Electromagnetic compatibility (EMC)

    International:

    • IEC 61326-1: Basic electromagnetic environment
    • CISPR 11: Group 1, Class A
      • Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself.
      • Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances.

        Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object.

    Korea (KCC):

    • Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes.

    USA (FCC):

    • 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.

    Electrical specifications

    Low resistance source
    Range 100 mΩ to 10 kΩ + 10 mΩ single value selection, dc and line frequency (50/60 Hz)

     

    1.5 kV high resistance source (DC only)
    Range 10 kΩ to 10 GΩ + 100 GΩ single value selection
    5.5 kV high resistance source (DC only) (5322A with /5 option)
    Range 10 kΩ to 100 GΩ
    Test voltage measurement
    Range 0 V dc to 5500 V dc
    Ground bond resistance source
    Resistance mode
    Range 1 mΩ to 1700 Ω, dc and line frequency (50/60 Hz)
    Line/loop impedance source
    Range 25 mΩ to 1700 Ω
    Correction manual/scan mode
    Residual impedance range 0 Ω to 10 Ω
    Resolution 1 mΩ
    Correction COMP mode (active loop compensation) (5322AVLC Option)
    Maximum compensated impedance 0 Ω to 2 Ω
    Leakage current source
    Range 0.1 to 30 mA
    Test voltage

    Passive mode: 60 V ac to 250 V ac rms

    Differential mode: 60 V ac to 250 V ac rms

    Substitute mode:10 V ac to 250 V ac rms

    Active mode (5322A/VLC only)[1]: 50 V ac to 100 V ac rms

    Uncertainty

    Passive mode: ± (0.3% setting + 2 μA)

    Differential mode: ± (0.3% setting + 2 μA)

    Test uncertainty can be influenced by power line voltage instability

    Substitute mode: ± (0.3% setting + 2 μA)

    Active mode (5322A/VLC only)[1]: ± (0.3% setting + 1 μA)

     

    [1] The Active Mode outputs are synchronized with the ac mains frequency to suppress interference between the calibrator and external noise sources.

    Human body simulation (for substitute leakage currently only

    Resistance range: 0 Ω to 10 000 Ω

    Resolution: 1 Ω

    RCD (residual current device) (for installation testers)
    Trip current range

    0.5 X I and 1 X I Mode: 5 to 30 mA in 1 mA steps

    1.4 X I and 2 X I Mode: 14 to 60 mA in 1 mA steps

    5 X I Mode: 50 to 150 mA in 1 mA steps

    Trip current measurement resolution

    1 μA below 30 mA

    10 μA in range from 30 mA to 150 mA

    100 μA in range from 300 mA to 3 A

    Trip current measurement uncertainty
    Trip current ± 1 % of nominal current (I) setting
    Trip time range 10 to 5000 ms
    Trip time uncertainty (0.02 % setting + 0.25 ms)
    Touch/line voltage

    Touch voltage range: 50 V

    Touch voltage setting: in discrete points depending on setup trip current value

    Touch series resistance: 0.02 Ω, 0.05 Ω, 0.10 Ω, 0.35 Ω, 0.50 Ω, 0.96 Ω, 1.7 Ω, 4.7 Ω, 9 Ω, 17 Ω, 47 Ω, 90 Ω, 170 Ω, 470 Ω, 900 Ω, 1700 Ω

    Line voltage range: 250 V

    Line voltage uncertainty: ± (5 % reading + 3 V)

    User selectable nominal line voltage: 100 V/115 V/120 V/220 V/230 V/240 V/250 V or real

    Post-trip delayed power restore mode: user selectable

    RCD (Residual current device (for PATs)
    Trip current range 0.5 X I to 5 X I Mode: 5 to 150 mA in 1 mA steps
    Trip current measurement resolution 1 μA
    Trip current measurement uncertainty Trip current: ± 1 % of nominal current (I) setting
    Trip time range 10 to 5000 ms
    Trip time uncertainty (0.02 % setting + 0.25 ms)
    Line voltage Reconnection delay: 2.5 s
    AC/DC voltage calibrator (5322A with VLC Option)
    Range 0.03 V to 600 V, ac or dc (.5 - 1%)
    Internal ranges

    AC mode: 0.3 V to 600 V

    DC mode: 0.3 V to 600 V

    Multimeter 
    Maximum withstand voltage

    HV terminal to COM terminal: 5000 V rms

    V terminal to COM terminal:  1100 V rms

    COM terminal to Protective Earth: 2200 V pk

    AC/DC voltage range

    V (1100 V) Input:  0 V dc to ±1100 V dc 10 mV to 1100 V ac rms

    HV (5000 V) Input: 0 Vdc to ±5000 Vdc 5 V to 5000 V ac rms

    AC/DC current
    Range 0 A to 20 A continuous, 20 A to 30 A for up to 5 minutes, ac rms or dc
    AC power
    Range 0 kVA ac to 33 kVA ac
    Type Apparent, active, reactive
    Phase indication Phase angle (φ), Power Factor (PF)
    Phase uncertainty (dφ) ± 0.1 °
    DC power
    Range 0 to 33 kVA dc
    Hipot ac voltage distortion measurement
    Frequency range 45 Hz to 65 Hz
    Number of harmonics 25
    Hipot dc voltage ripple coefficient measurement
    Voltage range 100 V dc to 5000 V dc
    Ripple coefficient range 10 %
    Flash test voltage measurement (using flash LC or flash V mode)
    Class I Voltage range 2000 V ac rms
    Uncertainty ± (0.3 % of reading + 6 V)
    Class II Voltage range 3000 V ac rms
    Uncertainty ± (1 % of reading value + 6 V)
    Flash leakage current measurement (using flash LC mode)
    Range 0 mA ac rms or dc to 300 mA ac rms or dc
    Resolution 4.5 digits

     